Autoclave Spore Test Log Sheet Guide for Dental Offices
Set up a dental spore test log, link biological indicator reports to loads, and work through an eight-row example across two sterilizers.
Set up the load and spore test records
Start with two records that share a sterilizer ID and can be linked by load ID. The load record captures the unit, cycle date and time, load ID, and number of packages. Include the mechanical readings or a reference to the cycle printout, along with the chemical monitoring observations required by your practice’s protocol. Record the actual cycle time, temperature, and pressure somewhere the reviewer can retrieve them.
The biological indicator (BI) record captures the unit, test date, associated load, recorded result and result date, and the lab report or in-office reader record reference. Track whether the source result has been retrieved and matched. Keep package counts on the load record so a second entry about a test does not increase the package total.
The CDC summary recommends biological, mechanical, and chemical monitoring together. The 2003 CDC guidelines describe at least weekly biological monitoring for each sterilizer and BI monitoring for each load containing an implantable device. Your practice’s monitoring schedule and log fields should follow its protocol and applicable requirements.
Reconcile eight rows across two sterilizers
In this illustrative review, the reviewer asks for Unit A loads dated November 9–15, 2026, inclusive. The export includes these eight rows.
| Cycle date | Unit | Load ID | Packages | Review scope |
|---|---|---|---|---|
| 2026-11-08 | Unit A | A-108 | 12 | Before window |
| 2026-11-09 | Unit A | A-109 | 22 | In window |
| 2026-11-10 | Unit B | B-110 | 14 | Other unit |
| 2026-11-11 | Unit A | A-111 | 18 | In window |
| 2026-11-12 | Unit B | B-112 | 16 | Other unit |
| 2026-11-13 | Unit A | A-113 | 10 | In window |
| 2026-11-14 | Unit B | B-114 | 11 | Other unit |
| 2026-11-16 | Unit A | A-116 | 20 | After window |
Filter on both unit and cycle date. The three in-window loads are A-109, A-111, and A-113. The total is 22 + 18 + 10 = 50 packages. Record the three Unit B rows and the two out-of-window Unit A rows as exclusions so a reviewer can reproduce the total from the same export.
Record a missing report reference
The BI entry tied to A-111 has a recorded result and an empty lab report reference. Log one follow-up to retrieve and match that report. The count is 1 missing lab reference = 1 open evidence item. Keep the 18 packages from A-111 in the review total because that load remains in scope.
Record who is retrieving the source and when it is due. Once the report is located, verify its unit and test date against the BI entry and record the reference and review date. Keep the original entry or its correction history available. The designated reviewer separately records any clinical disposition in the practice’s process.
Frequently asked questions
What fields belong on an autoclave spore test log sheet?
Record the sterilizer ID, test date, associated load ID, reported result and result date, and the lab or in-office reader record reference. Keep a follow-up status for source records awaiting verification.
Does a missing report reference change the package total?
Keep each in-scope load in the package total and track the missing report reference as a separate evidence item. In the example, three loads total 50 packages and one report reference is open.
CDC references
- CDC Summary of Infection Prevention Practices in Dental Settings — Sterilization and Disinfection (accessed 2026-11-28)
- CDC Guidelines for Infection Control in Dental Health-Care Settings (accessed 2026-11-28)